Overview
The OMX SR is a live cell three-dimensional (3D) structured illumination, super-resolution microscope (3D-SIM).
The system is unique because it enables up to three colours to be captured with two-fold better resolution than conventional microscopy and with high speeds compared to other super-resolution microscopy techniques. The platform also enables two-dimensional SIM and TIRF-SIM (total internal reflection fluorescence structured illumination microscopy). It is also suitable for fixed imaging.
This technique works best on cells or thin specimens (< 15 microns), that are brightly stained.
The system is compatible with standard slides, 35 mm dishes and 8-well chamber slides. It contains all necessary accessories for environmental control, for example CO2, humidity, and temperature control.
Staff at the Centre for Dynamic Imaging can provide expert guidance on the preparation of samples and choice of fluorophores.
Hardware
Magnification | 60x (Olympus) | 60x (Olympus) |
NA | 1.42 | 1.3 |
Immersion media | Oil | Silicone oil |
XY-resolution (µm) (@ 488 nm) 3D SIM |
120 nm | NA |
2D SIM | 120 nm | NA |
TIRF-SIM | 120 nm | NA |
WF Deconvolution | 220 nm | 240 nm |
Z-resolution (µm) (@ 488 nm) 3D SIM |
340 nm | NA |
2D SIM | 600 nm | NA |
TIRF-SIM | ~ 200 nm | NA |
WF Deconvolution | 600 nm | 600 nm |
Working distance (mm) | 0.15 | 0.3 |
Light sources
- 405 nm, 488 nm, 568, 640 nm
- LED trans-illuminator
Cameras
- 3 x PCO.edge sCMOS cameras
Technological Specifications (capabilities)
Available modalities
- 3D SIM (structural illumination microscopy)
- 2D SIM
- TIRF-SIM
- Very fast and multi-colour simultaneous widefield deconvolution
- Z-stack
- Time-series
- Tile scan
- Multi positions
- Live cell imaging
Additional hardware
- Blaze SIM Module OPTION for DeltaVision OMX SR
- Ultimate Focus Hardware Autofocus Module
- DIC optics
Unique features
- The OMX SR is a stable and compact microscope platform, which enables the dynamic capture of super-resolved images from live cells.
- The system enables fast three-colour live cell 3D-SIM imaging.
- A 60x/1.3 Silicone Oil objective is also available for live cell widefield deconvolution microscopy. Silicone Oil has a refractive index more similar to water and remains constant during long-term observations.
Limitations
- 9 x 15 images need to be captured at every focal plane for 3D-SIM, which means that photobleaching can be difficult.
- Weak staining, excessive bleaching or high levels of background can cause reconstruction artefacts.
Resources
Staff in the Centre for Dynamic Imaging can provide documentation to assist with the preparation of samples.